For optimal performance the experiment’s SQUID magnetometers must reside in a hermetically sealed Faraday cage. The aluminum Cryovessel serves this function, but temperature sensors and other devices with wiring that penetrates the Cryovessel must terminate in a connected Faraday cage (the Radio-Frequency Screen Room, or RFSR) outside the Magnetic Shield Enclosure. RFSR installation was completed, and the acceptance test passed.